  • 簡要參數:
    血液透析機質量檢測儀,全參數測量,電導率范圍0~199.9 mS/cm(電導探頭采用碳素現代4極設計),溫度范圍0~100 °C(分辨率0.01 °C),壓力范圍-700~+1900 mmHg(分辨率0.1mmHg),透析液流量范圍100~2000 ml/min (分辨率1 ml/min ),pH值范圍0~14pH(分辨率0.01pH),電壓范圍40VAC/DC(分辨率0.01V),頻率范圍250kHz(分辨率1Hz),時間最大1000sec(分辨率1msec),事件標記最大99999(分辨率1),流速模式動態測量或反滲水浸蘸模式測量,大屏幕,可觀察實時波形,內置可充電電池、交直流兩用,USB、RS232接口。
  • 人    氣:1986

相關推送

產品介紹

Dialysis    requires reliability and accuracy
   

The HDM99XP Meter allows the user to obtain reliable accuracy in measurement of      conductivity, temperature, pressure, flow, pH and voltage. Frequency and      period measurement are also possible. Developed by specialists with over 30 years of experience in the design of reference instruments for dialysis. The HDM99XP is an intelligent product featuring innovative technology, the end result is a product with an excellent price/performance ratio.
      

Advanced technology

Advanced processor      technology combined tightly with highly developed software results in a completely      new type of measuring instrument. Special features bring a level of      accuracy not achievable with previous units. An easy-to-read graphic      display and large keys make the HDM99XP convenient, not only to use, but handle. Operation       is simple and completely menu driven. The language displayed is selectable from the options menu, choices available are Dutch, English, French, German, Italian, Portuguese and Spanish.

There are  different ways of displaying measurements. These can be shown in graph format, including plotting. The graphs are adjusted automatically according to measuring range and  the time interval can be user selected. The progress of all  measurement parameters can be displayed at any time, this can greatly aid fault finding and diagnosis. There is an oscilloscope function built into the plotting  display for voltage measurement. Conductivity measurement is carried  out in 4 ranges up to 199.99 mS/cm, measuring ranges are automatically updated. The temperature reading is compensated in microsiemens or millisiemens, depending on which is selected.

Quality

      The HDM99XP is produced under certified quality management system no. ISO13485 The device conforms to CE requirements. Each unit will have passed rigorous inspections before delivery.  We also offer an annual calibration service and can provide a calibration certificate too upon request.

Trouble free    calibration

     Any operator can easily calibrate the unit via its intelligent menu-driven    software.     The software provides step by step menus aiding in the calibration process. To calibrate temperature, the temperature probe of the HDM is dipped into a liquid with a known temperature, this temperature can then be entered into the calibration program via the keyboard.  To increase the accuracy, up to six calibration values can be entered. All measurement parameters are calibrated in this way.

Technical Data

Conductivity
  
4 measuring ranges - Quadropole        electrode
       Value for temperature compensation selectable                      Range:0                to 199.9 mS/cm        Accuracy:0                to 199.9 uS/cm ± 0.6 uS/cm                200 to 1999 uS/cm ± 6 uS/cm 2 to 24.99 mS/cm ± 0.03 mS/cm 25 to 99.99 mS/cm ± 0.06 mS/cm > to 100 mS/cm ± 0.6 mS/cm
Temperature  Sensor integrated          in the conductivity electrode                      Range:0                to 100 °C        Resolution: 0.01 °C        Accuracy:25 ... 40°C ± 0,05 °C otherwise ± 0,07 °C
PressureReading selectable          as mmHg, kPa, atm, psi or bar                      Range:-700 to  +1900 mmHg, 2,5 Bar        Resolution:  0.1 mmHg        Accuracy:0                to 300 mmHg, ± 0.5 mmHg                otherwise ± 1 mmHg                in temperature range 20 to 23 °C
pH       Temperature via temperature probe in conductivity cell
         pH probe is optional                      Range:0                to 14                pH        Resolution: 0.01 pH        Accuracy:±                 0.02 pH
Flow    Range: 100 ... 2000 ml/min         Resolution:  1 ml/min        Accuracy:  100 ... 500 ml/min ±                 1.5 % otherwise ±                 2.5 %
Voltage  Range:40 V AC/DC,                up to 3kHz         Resolution: 0.01 V           Accuracy:± 1,0% of full measuring range value   
Frequency    Range: 250 kHzInput: 5                to 24 V        Resolution: 1 Hz        Accuracy:±                 0.2% of full measuring range value
Period Duration    Range: max. 1000 secInput: 5                to 24 V        Resolution: 1 msec        Accuracy:±                 0.2% of full measuring range value
Pulse/Event     Range: max. 99999Input: 5                to 24 V        Resolution: 1
Power          Supply  NiMH Battery rechargeable with 12 V DC          (charger included)
       Operation time up to 20 hours
Dimensions  270 x 120 x 50 mm  -        8" x 4" x          1.8" 
Weight    1.5 Kg or 3.3 lbs          including Conductivity Probe

Connector